Recall Z-1484-2018 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ven

DateMay 8, 2018 (initiated Mar 1, 2018)
StatusTerminated
ReasonThe automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Root causeProcess control
ActionOn March 1, 2018, the recalling firm notified affected customers via an Urgent Medical Device Recall letter. The recalling firm stated that customers would receive replacement motors for potentially affected devices. Until the replacement occurred, customers were instructed to continue to operate their devices with the usual attention and make sure that a manual resuscitator for emergency ventilation is kept ready according to the instructions for use. Customers were also supplied with a reply card to indicate that they had received the notification. Customers with questions about the letter
Quantity1
DistributionUS Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042419FABIUS GS/FABIUS TIRO ANESTHESIA SYSTEM
Draeger Medical, Inc., Oct 8, 2004