ClearanceBrief
Recall Z-1399-2018 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ONCOR Expression, Digital Linear Accelerator, Model No. 07360717 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

DateFeb 14, 2018
StatusTerminated
ReasonControl Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Root causeSoftware design
ActionOn February 16, 2018, the recalling firm issued a Customer Advisory Notice via USPS Certified Mail to inform customers about the update of the Control Console software version 9.2.400 of the Digital Linear Accelerator to version 9.2.502. Customers were informed of the issues that would be fixed through this software update. Consignees were asked to instruct affected personnel accordingly, place the safety advisory notice in the device's instructions for use, and have personnel maintain awareness until the fix is implemented.
Quantity2
DistributionWorldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006