Recall Z-1596-2018 ยท FFK Lithotriptor, Electro-Hydraulic
Cook Inc.
StoneBreaker Single Use Probe, Catalog numbers: SBP-010500; SBP-010605; SBP-016500; SBP-016605; SBP-020425 GPN : G52607; G52605; G52606; G52609; G52608
| Date | Jan 31, 2018 |
| Status | Terminated |
| Reason | The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use. |
| Root cause | Error in labeling |
| Action | On March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secon |
| Quantity | 12,556 |
| Distribution | US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, U |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K062475 | LMA STONEBREAKER DEVICE; PROBE SIZE 1.0MM, 1.6MM, 2.0MM The Laryngeal Mask Co., Ltd., May 9, 2007 |