ClearanceBrief
Recall Z-1593-2018 ยท FFK Lithotriptor, Electro-Hydraulic

Cook Inc.

StoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013

DateJan 31, 2018
StatusTerminated
ReasonThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Root causeError in labeling
ActionOn March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secon
Quantity370
DistributionUS Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, U
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062475LMA STONEBREAKER DEVICE; PROBE SIZE 1.0MM, 1.6MM, 2.0MM
The Laryngeal Mask Co., Ltd., May 9, 2007