ClearanceBrief
Recall Z-1378-2018 ยท MQG Accessory, Assisted Reproduction

Cook Inc.

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

DateApr 13, 2018 (initiated Jan 12, 2018)
StatusTerminated
ReasonThe temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Root causeDevice Design
ActionCook plans to begin notifying consignees of the recall on January 12, 2018 Written recall communication letters will be mailed via courier to US consignees
Quantity4,451
DistributionNationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983642COOK MINI-INCUBATOR
Cook Ob/Gyn, Feb 1, 1999