Recall Z-1395-2018 ยท KGI Densitometer, Bone

GE Medical Systems Ultrasound & Primary Care Diagnostics, LL

GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734, LU42344, LU42365 b) Prodigy Advanced, Model Numbers: LU42361, LU42397 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discre

DateMar 16, 2018
StatusTerminated
ReasonUnder certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Root causeSoftware design
ActionThe firm initiated their recall by letter on March 16, 2018. The notice stated: "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS. There have been no injuries reported as a result of this issue. You can continue using your system by disabling the DICOM MPPS feature as follows: 1. In enCORE, select Tools | Co
Quantity303 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982267DPX-RX BONE DENSITOMETER
Lunar Corp., Aug 18, 1998