Recall Z-1393-2018 ยท KGI Densitometer, Bone

GE Medical Systems Ultrasound & Primary Care Diagnostics, LL

GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.

DateMar 16, 2018
StatusTerminated
ReasonUnder certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Root causeSoftware design
ActionThe firm initiated their recall by letter on March 16, 2018. The notice stated: "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS. There have been no injuries reported as a result of this issue. You can continue using your system by disabling the DICOM MPPS feature as follows: 1. In enCORE, select Tools | Co
Quantity26 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030962DPX SERIES BRAVO DUO BONE DENSITOMETER
Ge Lunar Corp., Jul 25, 2003