Recall Z-1546-2018 ยท GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered

Cardinal Health 200, LLC

Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body fluids. These fluids may be suctioned during surgical procedures or other applications requiring the removal of built up fluids. The suction canisters are used in operating rooms but are also used outside the surgical field as well as post-operat

DateMar 28, 2018
StatusTerminated
Reasonproducts may be missing a one-way valve on the fluid patient port.
Root causeProcess control
ActionCardinal Health sent an Urgent Product Recall letter dated March 28, 2018, Please take the following actions: 1. INSPECT your inventory for the affected product code and lot number (listed on package) 2. SEGREGATE and QUARANTINE all on-hand product 3. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652- 9648, whether or not you have affected product, indicating the product code, lot and quantity of product you've quarantined 4. CONTACT Customer Service at 800-964-5227 to arrange for return and credit of any affected product . For further questions, please call (800)
Quantity3600 units
DistributionUS Distribution to the state of MA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.