Recall Z-1611-2018 ยท CEM Electrode, Ion Specific, Potassium

Roche Diagnostics Corporation

Cobas b 221<4>Roche OMNI S4 system; Catalog numbers: 3337138001 and 3337138692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and

DateMay 3, 2018 (initiated Feb 22, 2018)
StatusTerminated
ReasonThe software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Root causeSoftware design
ActionROCHE sent an Urgent Medical Device Correction Letter dated February 22, 2018, via UPS to affected customers. Customers were instructed to complete the following: " 1) Use the Return to Setup and Exit buttons to ensure execution of scheduled AutoQC measurements as described in this Urgent Medical Device Correction (UMDC); 2) Check to see if the next scheduled AutoQC measurement is performed to ensure the AutoQC scheduler is working as expected; 3) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site; 4) Complete the attache
Quantity1026 total products
DistributionUSA ( nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032311ROCHE DIAGNOSTICS OMNI S ANALYZER
Roche Diagnostics Corp., Oct 17, 2003