ClearanceBrief
Recall Z-1537-2018 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer Biomet, Inc.

Modular Head Component, Biomet Hip System; Item No. 163669, Hip Joint, metal/polymer semi-constrained cemented prosthesis The device is used in treatment of patients who need total hip arthroplasty, and consists of modular components and corresponding instruments.

DateApr 2, 2018
StatusTerminated
ReasonTwo lots of different sized modular heads potentially commingled. Risks include delay in surgery and dislocation due to use of an incorrect sized head.
Root causeNonconforming Material/Component
ActionThe firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 4/2/2018 to its customers. The letter described the product, problem and actions to be taken. Distributors will be notified of the recall via email. Hospital risk managers and distributors with product will be notified via courier. The distributor letter includes instructions to locate and remove the product in their territory, identify hospitals who have previously used the product; return on-hand product to the recalling firm and ensure all products are accounted for. The letter for hospital risk managers instruct t
Quantity30
DistributionWorldwide Distribution: US (nationwide) in states of: AL, AR, FL, GA, IN, KY, LA, MO, MS, OH, OK, TX, and VA; and countries of: Canada, Costa Rica and Chile.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911684BIOMET COBALT-CHROME FEMORAL COMPONENTS
Biomet, Inc., Jun 27, 1991