Recall Z-1587-2018 ยท JGS Electrode, Ion Specific, Sodium

Siemens Healthcare Diagnostics, Inc.

Atellica Solution and Atellica Sample Handler Prime for in vitro diagnostic testing of clinical specimens.

DateFeb 26, 2018
StatusTerminated
ReasonWhen more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
Root causeSoftware design
ActionRecall initiated 02/23/2018 Firm released an update software (SW) version (V), 1.12.1, which corrects the behavior. A Mandatory Update (MU) campaign was launched to install the new SW V on all systems, whether or not they are impacted by this behavior. The MU campaign will be tracked to 100% completion globally.
DistributionFL and France Spain Sweden United Kingdom Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161954Atellica Solution, Atellica A-LYTE Integrated Multisensor Na+/K+/CI-, Atellica CH Albumin BCP Reagent (Alb_p), Atellica
Siemens Healthcare Diagnostics, Mar 9, 2017