Recall Z-1412-2018 ยท JAK System, X-Ray, Tomography, Computed
Philips Medical Systems (Cleveland) Inc
Brilliance 16 Slice (Air) - Model no. 728246, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
| Date | Feb 8, 2018 |
| Status | Terminated |
| Reason | Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities. |
| Root cause | Software design |
| Action | The recalling firm notified customers of the recall on February 8, 2018, via Urgent Field Safety Notice. The notice advised customers of the issue and affected products and lot numbers. Customers were advised that once the E-stop has been closed, the operator/technologist must initate a brief horizontal motion (in or out) using any of the operator control panels, CT-Box or tape/foot switches. This action will ensure the couch carbon top (upper pallet) is in a controlled state. Customers were informed that a Field Service Engineer would contact them to schedule site updates to correct the issue |
| Quantity | 2 |
| Distribution | Worldwide Distribution - U.S. Nationwide. International distribution to Canada, Mexico, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cuba, Czech Republic, Denmark, Djibouti, Dominican Republ |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012009 | MX8000 IDT CT SYSTEM Marconi Medical Systems, Inc., Sep 13, 2001 |