Recall Z-1524-2018 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Zimmer GmbH

Avenir Muller Stem; Item Number: 01.06010.003, Lot Number: 2904368, UDI: (01) 00889024479494 (17) 220331 (10) 2904368

DateFeb 28, 2018
StatusTerminated
ReasonA potential commingle that could result in the product in the package not matching the product on the outer label.
Root causeError in labeling
ActionOn March 7, 2018 Field Safety Corrective Action Notices were hand delivered by Zimmer Biomet team members or their representatives. Customers were instructed to complete the following: 1.Review this notification for awareness of the contents. 2.Assist your Zimmer Biomet sales representative to quarantine immediately all affected implants. 3.Your Zimmer Biomet sales representative will remove the affected implants from your facility. 4.Complete Attachment 1 Certificate of Acknowledgement. a. Return a digital copy to fieldaction.emea@zimmerbiomet.com. b. Retain a copy of the Certificate of Ac
DistributionThere was no distribution to the US. The distribution occurred to the following countries: Belgium, Switzerland, Germany, France, Canary Islands, Italy, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123392AVENIR MULLER STEM
Zimmer GmbH, Mar 4, 2013