Recall Z-1416-2018 ยท DWC Controller, Temperature, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corporation
Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.
| Date | Apr 18, 2018 (initiated Mar 16, 2018) |
| Status | Terminated |
| Reason | Update to cooler-heater cleaning instructions. |
| Root cause | Labeling design |
| Action | The firm,Terumo Cardiovascular Systems Corporation, mailed customers an "Urgent Medical Device Correction" letter dated 2/16/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) discard the current Sarns TCM, TCM II Cooling and Heating, or tHX2 Temperature Management Systems cleaning guides; 2) Review the notice and updated cleaning guides, and ensure all users receive notice of the updated cleaning guides and; 3) Return the Customer Response Form acknowledging receipt of the notice via Terumo Recall Email: tcvs.recall@t |
| Quantity | 2393 this total is for all products |
| Distribution | Worldwide Distribution: US (Nationwide), Australia, Belgium, Canada, Chile, China. Colombia, Dominican Republic, England, Finland, France, Germany, Greece, HK, India, Indonesia, Iran. Israel, Italy, Japan, Korea, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K071521 | TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM Terumo Cardiovascular Systems Corp., Nov 21, 2007 |