Recall Z-1480-2018 ยท PQP Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System

Life Technologies Corporation

Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA/DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test.

DateFeb 8, 2018
StatusTerminated
ReasonThe lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.
Root causeProcess control
ActionThe firm issued an Urgent: Medical Device Recall Notification, dated February 16, 2018, via e-mail and phone call to affected consignees. The letter identified the affected device and advised customers of the potential "No Call" or false positive outcomes. Users were advised that they may need an alternate method to determine the presence of ROS1 variant in previously tested patient samples, and that any future tests using the affected lot cannot be considered valid if "No Call" or "Negative" is the result. Requested action included reviewing previous results which used the affected lot.
Quantity30
DistributionDistributed to states: AZ, CA, NC, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.