Recall Z-1383-2018 ยท DSP System, Balloon, Intra-Aortic And Control
Datascope Corporation
Intra-Aortic Balloon Catheter Mega 7.5FR 40cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
| Date | Apr 16, 2018 (initiated Feb 21, 2018) |
| Status | Terminated |
| Reason | The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section. |
| Root cause | Error in labeling |
| Action | Maquet/Getinge issued a recall letter on February 21, 2018 to affected customers, via FEDEX with confirmation of delivery, to notify them of the fact that the IFU is missing the Contraindications section and to provide the missing information. Affected customers are instructed to examine their inventory immediately to determine if they have affected product. An Addendum to the Mega IAB Catheter IFU, containing the Contraindications, is provided as an attachment in the recall letter. It is suggested that affected customers post copies of the Addendum near all affected lot numbers of Mega IA |
| Quantity | 2714 |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120868 | MEGA(TM) 7.5FR. 30CC INTRA-AORTIC BALLOON CATHETER AND ACCESSORIES, MEGA(TM) 7.5FR. 40CC INTRA-AORTIC BALLOON CATHETER Datascope Corp., Cardiac Assist Division, Apr 20, 2012 |