Recall Z-1379-2018 ยท MEC Disinfector, Medical Devices

Getinge Disinfection Ab

Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 86 Series UDI Info: 86 Series Washer Disinfector 07340153700246, 07340153700253, 07340153700260, 07340153700277 Getinge automated loading/unloading systems are intended to be used as an automatic conveyor/transport system for unloading carriers (wash carts) from Getinge 86/88 series Washer-disinfectors. The automated loading/unloadin

DateMar 2, 2018
StatusTerminated
ReasonThere is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
Root causeSoftware design
ActionA letter (Urgent Medical Device Recall Notice) was issued March 2, 2018 via 2 Day Fedex with customer signature required to allow for tracking of delivery. The letter includes a customer response form requesting feedback including if device is still in use at the facility. The letter instructs the customer to contact Getinge Service to schedule a visit and service by a trained Getinge technician at no charge to the customer. In addition Getinge Service will proactively contact affected customers and perform the correction when affected installed devices are identified. Recalled product will b
Quantity170 total products
DistributionWorldwide Distribution - USA (nationwide) and the following countries: AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, NETHERLANDS, NORWAY, SPAIN, FINLAND, FRANCE, GREAT BRITAIN, HONG KONG, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NORWAY, NEW ZELAND, SAUDI ARAB
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.