Recall Z-1487-2018 ยท ITG Bandage, Cast

Integra LifeSciences Corp.

TCC-EZ, Total Contact Cast System, Model No. TCC2-4051

DateAug 18, 2017
StatusTerminated
ReasonA quarantined lot of the devices was released for sale. Devices from this lot may experience increased exothermic reaction causing higher levels of heat output which exceed the product specifications.
Root causeProcess control
ActionOn 8/18/17, the recalling firm's sales reps notified affected consignees of the recall and distributed customer letters. Customers were asked to examine their inventory for product from the affected lot, warned not to use the product, complete the included "Acknowledgement and Return Form", and return the form by email or fax. Customers were provided with a Customer Service number: 1-800-654-2873.
Quantity250
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.