Recall Z-1429-2018 ยท BSY Catheters, Suction, Tracheobronchial
Halyard Health, Inc
Halyard Closed Suction System for Adults, Multi-AccessPortClosedSuctionSystemforAdults. Used to aspirate liquids or semisolids from a patient's upper airway. a. HALYARD* MULTI-PORT, DSY, ENDO 10FR; b. HALYARD* MULTI-PORT, DSY, ENDO 12FR; c. HALYARD* MULTI-PORT, DSY, ENDO, MDI 12FR; d. HALYARD* MULTI-PORT, DSY, ENDO 14FR; e. HALYARD* MULTI-PORT, DSY, TRACH 14FR; f. HALYARD* MULTI-PORT
| Date | Feb 7, 2018 |
| Status | Terminated |
| Reason | Certain Flex Connectors supplied with Halyard Closed Suction Kits with Flex Connector may become loose or disconnect before use or during use. |
| Root cause | Labeling design |
| Action | Customers were notified via mail and/or email on about 02/07/2018. Instructions for distributors included to notify all affected customers who may have received potentially impacted products and to complete and return the Distributor Acknowledgement Form. Instructions for customers included to ensure adequate connection of the Flex Connector to Swivel Connector before use of the product by following. This accomplished by ensuring a tight fit is achieved on the Swivel with the Flex Connector advanced approximately three-quarters over the Swivel, so a gap remains visible. If and adequate c |
| Quantity | 32,189 cases (20 units per case - 643,780 units) total |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, ARGENTINA, AUSTRALIA, BAHAMAS, BELGIUM, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., ECUADOR, EL SALVADOR, GUATEMALA, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JA |
| FDA record | Open on fda.gov |