Recall Z-1478-2018 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system.

DateJan 19, 2018
StatusTerminated
ReasonInaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).
Root causeEmployee error
ActionThe firm notified customers via phone alerting them of this affected device, issue, potential hazard, correction, and instructions. The communication also included scheduling an expedited service call for a Field Service Technician to replace the affected Flowmeter Module with a corrected Flowmeter Module. When necessary to avoid delaying or cancelling life-sustaining surgery, users can continue to use the Flowmeter Module while awaiting replacement. Once the issue is recognized by the user, if a replacement Flowmeter Module is available, replacement and reassignment of safety connections
Quantity7
DistributionWorldwide Distribution -- USA, including the states of IL, KY, and MI; and, the country of Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172220Advanced Perfusion System 1
Terumo Cardiovascular Systems Corporation, Aug 17, 2017