Recall Z-1478-2018 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corporation
Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system.
| Date | Jan 19, 2018 |
| Status | Terminated |
| Reason | Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation). |
| Root cause | Employee error |
| Action | The firm notified customers via phone alerting them of this affected device, issue, potential hazard, correction, and instructions. The communication also included scheduling an expedited service call for a Field Service Technician to replace the affected Flowmeter Module with a corrected Flowmeter Module. When necessary to avoid delaying or cancelling life-sustaining surgery, users can continue to use the Flowmeter Module while awaiting replacement. Once the issue is recognized by the user, if a replacement Flowmeter Module is available, replacement and reassignment of safety connections |
| Quantity | 7 |
| Distribution | Worldwide Distribution -- USA, including the states of IL, KY, and MI; and, the country of Singapore. |
| FDA record | Open on fda.gov |