Recall Z-1388-2018 ยท GHQ Stain, Fetal Hemoglobin

Life Technologies Corporation

Human Fetal Hemoglobin (HBF-1) FITC Conjugate

DateFeb 7, 2018
StatusTerminated
ReasonThree lots of Human Fetal Hemoglobin FITC Conjugate are displaying high background and unusual bright staining in the controls and patient samples which could potentially lead to erroneous results (i.e. false positives)
Root causeUse error
ActionAll consignees which have purchased the Human Fetal Hemoglobin FITC antibody including facilities/customers that have shipped the lots outside of their facility will be notified via certified mail to check inventory to see if affected product is on hand. Consignees instructed to discontinue use and discard per internal procedure. The effectiveness check will begin 10 days after the distribution date of the Medical Device Recall Notification. The field removal is considered effective when 10% of the affected customers acknowledge receipt and understanding of the notification instructions.
DistributionMI, CA, IL, TX, SC, NY, MN, DE, NJ, WA , OR, IN, MA, DE, FL, PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990641CALTAG FETAL HEMOGLOBIN TEST
Caltag Laboratories, Inc., Sep 17, 1999