Recall Z-1867-2018 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Invivo Corporation

PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers: 453564241901, 453564483321, 453564621791, and 989803199561 (international only) The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health c

DateMar 14, 2018
StatusTerminated
ReasonThe real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
Root causeSoftware design
ActionThe firm initiated their recall by letter on March 14, 2018. The Field Safety Notice informed the consignee about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients " the actions planned by Philips to correct the problem. If further information or support concerning this problem is needed, the consignee was directed contact the local Philips representative: 800-669-1328 option 2, then option 3.
Quantity4375 units
DistributionWorldwide Distribution - USA (nationwide) Distribution including Puerto Rico., and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Kazakhstan, K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.