Recall Z-1318-2018 ยท HKN Refractor, Manual, Non-Powered, Including Phoropter

Reliance Medical Products Inc

Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position ophthalmological examination and diagnostic equipment, is to store and position for examination a device commonly known as a refractor.

DateDec 6, 2017
StatusTerminated
ReasonProduct has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing
Root causeEmployee error
ActionRecalling firm sent letter to customers via certified mail for US customers and via regular mail to foreign customers. Verify product has the presence of the one or more of the three parts: Refractor Suspension Arm missing Screw, Washer and/or Retaining Ring (Snap Ring) that may be missing, and to assure it is functioning properly. For further questions, please call (513) 398-3937
Quantity933 ( of which 97 has been quarantined)
DistributionWorldwide Distribution - USA(nationwide) Distribution to the states of : NJ, CA, OH, VA, MA, MN, IL , UT, TN, AR, LA, AZ, PA, TX, IA and HI., and to the countries of : Costa Rica, Jamaica, South Africa, New Zealand, South Korea, Dubai, Saudi Arabia and Jordan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.