Recall Z-1438-2018 ยท LNH System, Nuclear Magnetic Resonance Imaging
Philips Electronics North America Corporation
T5-NT, Model 78104 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
| Date | Apr 18, 2018 (initiated Mar 16, 2018) |
| Status | Terminated |
| Reason | Potential risk for helium gas inside the MR examination room during a magnet quench |
| Root cause | Nonconforming Material/Component |
| Action | On March 16, 2018, Philips Healthcare distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via certified mailing. On May 24, 2018, the recall letter was distributed to additional affected customers identified by the firm. **Action to be Taken by Customer**: During a quench, a large amount of helium evaporates and is vented outside the building through a venting system. A quench causes immediate removal of the static magnetic field. A quench can occur spontaneously or can be induced if there is an emergency. In case helium gas escapes into the examination room during a mag |
| Quantity | 8,205 units in total |
| Distribution | Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile, China, Colombia, Congo (Democratic Republic o |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K945945 | GYROSCAN ACS-NT, GYROSCAN T5-NT AND GYROSCAN T10-NT Philips Medical Systems North America, Inc., Dec 22, 1995 |