ClearanceBrief
Recall Z-1434-2018 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Electronics North America Corporation

Panorama 1.0T, Model 781250 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.

DateApr 18, 2018 (initiated Mar 16, 2018)
StatusTerminated
ReasonPotential risk for helium gas inside the MR examination room during a magnet quench
Root causeNonconforming Material/Component
ActionOn March 16, 2018, Philips Healthcare distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via certified mailing. On May 24, 2018, the recall letter was distributed to additional affected customers identified by the firm. **Action to be Taken by Customer**: During a quench, a large amount of helium evaporates and is vented outside the building through a venting system. A quench causes immediate removal of the static magnetic field. A quench can occur spontaneously or can be induced if there is an emergency. In case helium gas escapes into the examination room during a mag
Quantity8,205 units in total
DistributionWorldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile, China, Colombia, Congo (Democratic Republic o
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041602PANORAMA 1.0T
Philips Medical Systems North America, Inc., Jul 26, 2004