Recall Z-1349-2018 ยท FSY Light, Surgical, Ceiling Mounted

GETINGE US SALES LLC

VOLISTA StandOP Surgical Light; Model numbers VLT600 SF AIM STP, VLT600 DF AIM STP, Ref. codes ARD568811901, ARD568811911 ARD568811951, ARD568811961 The device is designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. These surgical lights may be installed in surgical suites, examining rooms, doctors' surgeries and outpatient consultations.

DateApr 10, 2018 (initiated Feb 7, 2018)
StatusTerminated
ReasonThe manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall into the sterile field during a procedure.
Root causeComponent design/selection
ActionThe firm, Getinge, sent an "URGENT MEDICAL DEVICE RECALL" Notice dated 2/7/2018 to all affected consignees via mail. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: 1) Ensure that all users of the device are made aware of this field correction. 2) Check the device serial number against the list of affected serial numbers. 3) Contact the recalling firm to arrange for servicing of all affected units. Getinge Customer Service: (800) 950-9912, 8AM - 5PM (EST) Monday to Friday. 4) Complete and sign the enclosed customer response
Quantity5214
DistributionWorldwide Distribution: US (nationwide) to states of: AR, AZ, CA, CO, FL, ID, IA, KS, KY, ME, MO, NE, NY, OR, TN, TX, UT, and WA; and countries of: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Est