Recall Z-1200-2018 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.

DateMar 26, 2018 (initiated Mar 5, 2018)
StatusTerminated
ReasonThe ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.
Root causeDevice Design
ActionUndated recall letters were issued via email (if they had an email address for the customer) or via certified mail (if no email address was available) on 3/5/2018.
Quantity229 systems
DistributionDistribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170315Acuson SC2000 Diagnostic Ultrasound System