Recall Z-1170-2018 ยท PMP Dorsal Root Ganglion Stimulator For Pain Relief

St. Jude Medical, Inc.

Proclaim DRG Implantable Pulse Generator, Model Number 3664

DateMar 23, 2018 (initiated Mar 9, 2018)
StatusTerminated
ReasonThe firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.
Root causeSoftware change control
ActionNotification material in the form of a Customer Letter was hand-delivered to two consignees in the US beginning 09 Mar 2018. The Physician Letter was hand-delivered to impacted implanting physicians, beginning 14 Mar 2018. A Patient Letter was delivered via certified mailing to all active implanted patients in the US, beginning 14 Mar 2018.
Quantity17 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.