Recall Z-1321-2018 ยท GIF Diluent, Blood Cell

Mindray DS USA, Inc. dba Mindray North America

Hematology Diluent M-53D used with BC-5390 Hematology Analyzer; Part Number: 105-007867-00 (20L container) The M-53D diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-5390 Hematology Analyzer.

DateJan 24, 2018
StatusTerminated
ReasonCertain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzer may cause the system to produce an elevated platelet background count.
Root causeMaterial/Component Contamination
ActionLetters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.
Quantity496
DistributionUS Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.