Recall Z-1337-2018 ยท EQJ Bur, Ear, Nose And Throat

Gyrus ACMI, Inc.

Gyrus ACMI Round Cutting Burrs - to be used with the Diego Elite system

DateMar 20, 2018 (initiated Aug 3, 2017)
StatusOpen, Classified
Reason"Chattering" when cutting burrs are used in bone.
ActionAs a result, OSTA will issue a Field Safety Notice (FSN) to all customers that received variable exposure burrs of 5mm or greater diameter. The FSN will inform them of the concern and will advise to exercise due caution if using the affected device in proximity to sensitive or vital anatomical structures. On September 6, 2017 the firm sent a URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE to the attention of the Surgical Risk Management Department for the following products: MBUR5060FRCV - Burr, 5mm Round Cutting, Variable Exposure, Standard Length, MBUR6060FRCV - Burr, 6mm Round Cutting, Variabl
Quantity324
DistributionNM, MI, and TX EU, Australia, Japan, and Hong Kong
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.