Recall Z-1138-2018 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

ESAOTE S.P.A.

MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33

DateJan 3, 2018
StatusTerminated
ReasonThe ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
Root causeSoftware design
ActionThe January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.
Quantity5
DistributionUS distribution in the states: AZ, FL, OK, PA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081386MODEL 6150 (MYLAB70) AND MODEL 6100 (MYLAB90)
Esaote, S.p.A., Sep 25, 2008
K1630826200 Ultrasound System and 6250 Ultrasound System
Esaote, S.p.A., Mar 16, 2017