Recall Z-1303-2018 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo.via. Medical Device Software. Picture archiving and communications system.

DateMar 20, 2018 (initiated Jan 8, 2018)
StatusTerminated
ReasonA software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.
Root causeSoftware design
ActionCustomer Advisory Notice SY065/17/P was sent to direct accounts. Firm is continuing to work on software solution. A workaround is available and is described in the Customer Advisory Notice.
Quantity236 units
DistributionDevice is software only. No products are distributed to wholesale dealers, distributers or retailers
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123375SYNGO.VIA
Siemens AG, Nov 20, 2012