Recall Z-1133-2018 ยท FBN Choledochoscope And Accessories, Flexible/Rigid

Boston Scientific Corporation

SpyScope DS Access and Delivery diagnostic biliary catheter, M00546600

DateDec 20, 2017
StatusTerminated
ReasonCertain lots of devices may have the working channel sleeve protruding from the camera cap which may result in injuries from minor tissue damage to significant bleeding.
Root causeProcess design
ActionThe notification letter requests the consignee to identify and return the affected product along with the response form. Distributors are asked to share the recall information to the customer level. 12/20/2017 by mail.
Quantity5026
DistributionWorldwide Distribution US nationwide, Austria, Belarus, Belguim, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hungary, Ireland, Israel, Italy, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Spain, Sweden, Switzerland, Bahrain, China, Hong Kong, India, Malaysia, Pakist
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142922SpyGlass DS Direct Visualization System
Boston Scientific Corporation, Jan 16, 2015