Recall Z-1329-2018 ยท DYB Introducer, Catheter

Arrow International Inc

(1) Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port, Sharps Safety Features, and Maximal Barrier Precautions for use with 7.5 - 8 Fr. Catheters, Part Number: ASK-09803-WBH1; (2) Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port, Sharps Safety Features, and Maximal Barrier Precautions for use with 7.5 - 8 Fr. Catheters, Part Number: ASK-09903-KH

DateDec 22, 2017
StatusTerminated
ReasonCertain lots of cardiovascular catheters may not be completely sealed and so sterility of the product cannot be guaranteed.
Root causeProcess control
ActionThe notification letter includes the following instructions for CUSTOMERS: 1. Discontinue use and quarantine any affected products. 2. Complete and return the Recall Acknowledgement Form to start return process. 3. Complete the Recall Acknowledgment Form if there is no product on hand. Instructions for DISTRIBUTORS: 1.Discontinue distribution and quarantine any affected products. 2. Communicate this recall to any customers with affected product. 3. Have the customers return any affected product to you with a completed Recall Acknowledgement Form. 4. Return affected products from inventory usin
DistributionDistributed in all 50 U.S. states and the District of Columbia. Distributed in Bahamas, Canada, Mexico, Panama, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K780532PERCUTANEOUS SHEATH INTRODUCER KIT
Arrow Intl., Inc., Apr 21, 1978