Recall Z-1319-2018 ยท FNL Bed, Ac-Powered Adjustable Hospital

Hill-Rom, Inc.

Progressa Bed - AC-powered adjustable hospital bed

DateJan 12, 2018
StatusTerminated
ReasonA component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.
Root causeComponent change control
ActionFirm has developed a correction for the braking system and plan to have a Hill-Rom technician to update affected Progressa Beds. Hill-Rom sent a letter, dated January 12, 2018. Until beds are updated, follow the safety information as outlined in the Progressa User manual each time when setting the brakes. Warning: Always set the brakes when the bed is occupied, except during patient transport. To help make sure the bed will not move, push and pull on the bed to check it after the brakes are engaged. Brakes should always be set when the bed is occupied and especially when moving a patie
Quantity2117
DistributionWorldwide distribution. US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.