ClearanceBrief
Recall Z-1691-2018 ยท MOI System, Test, Immunological, Antigen, Tumor

Siemens Healthcare Diagnostics, Inc.

IMMULITE 2000/IMMULITE 2000 XPi BR-MA (CA15-3)

DateMay 4, 2018 (initiated Jan 2, 2018)
StatusTerminated
ReasonConcentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Root causeError in labeling
ActionSiemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
DistributionNationwide and Canada, Mexico, and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013984IMMULITE BR-MA, MODEL #LKBR; IMMULITE 2000 BR-MA, MODEL #L2KBR
Diagnostic Products Corp., Mar 28, 2002