Recall Z-1322-2018 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Clinical Diagnostic Systems

VITROS Immunodiagnostic Products TSH Reagent Pack

DateApr 10, 2018 (initiated Jan 11, 2018)
StatusTerminated
ReasonVITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
Root causeNonconforming Material/Component
ActionDistributors instructed to disgard remaining inventory, complete Confirmation of Receipt form, and firm promised to replace product. Distributors instructed to send letter to customers.
QuantityLot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)
DistributionDistributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083173VITROS IMMUNODIAGNOSTIC PRODUCTS TSH REAGENT PACK AND CALIBRATORS, VITROS 3600 IMMUNODIAGNOSTIC SYSTEM
Ortho-Clinical Diagnostics, Dec 22, 2008