Recall Z-1688-2018 ยท JKC Radioimmunoassay, Dehydroepiandrosterone (Free And Sulfate)

Siemens Healthcare Diagnostics, Inc.

ADVIA Centaur DHEA-SO4

DateMay 4, 2018 (initiated Jan 2, 2018)
StatusTerminated
ReasonThe concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Root causeError in labeling
ActionSiemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA