Recall Z-1668-2018 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Siemens Healthcare Diagnostics, Inc.

Dimension Thyroid Stimulating Hormone (TSHL) Flex reagent cartridge

DateMay 4, 2018 (initiated Jan 2, 2018)
StatusTerminated
ReasonThe concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Root causeError in labeling
ActionSiemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
DistributionNationwide and Canada, Mexico, and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140842DIMENSION LOCI FREE THYROXINE FLEX REAGENT CARTRIDGE, FT4L, DIMENSION LOCI THYROID STIMULATING HORMONE FLEX REAGENT CART
Siemens Healthcare Diagnostics, Inc., Nov 26, 2014