ClearanceBrief
Recall Z-1363-2018 ยท NKG Posterior Cervical Screw System

Orthofix, Inc

ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003

DateFeb 23, 2018
StatusTerminated
ReasonOrthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Root causeNonconforming Material/Component
ActionThe firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
Quantity397 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172194Connector System
Orthofix, Inc., Aug 18, 2017