Recall Z-1266-2018 ยท FQO Table, Operating-Room, Ac-Powered

Deerfield Imaging, Inc.

IMRIS ORT300, Removable Operating Room Table, Part Numbers: (a) 114093-000 (b) 114093-600

DateFeb 23, 2018
StatusTerminated
ReasonThe Operating Room Table may drift while in use, which could potentially result in unintended movement of the table.
Root causeOther
ActionThe firm initiated their field notification on 02/23/2018 by letter. The notice stated the following: "It has recently come to our attention that the ORTT2X/100/200/300 tables may drift while in use, which could potentially result in unintended movement of the table. This may pose a risk of injury to the patient. To minimize these risks, IMRIS is taking the following corrective actions: " ORT T2X/100/200 tables: The tables will be tested for drift symptoms. No further action will occur for tables that pass the drift test. For tables that do not pass the drift test, a hydraulc fluid flush a
Quantity13 units
DistributionUS, Canada, China, Australia, Sweden, Germany