ClearanceBrief
Recall Z-1042-2018 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Biomet, Inc.

CR OPTION FEM COMP SIZE C/L CR OPTION FEM COMP SIZE C/R CR OPTION FEM COMP SIZE D/L CR OPTION FEM COMP SIZE D/R CR OPTION FEM COMP SIZE E/L CR OPTION FEM COMP SIZE E/R CR OPTION FEM COMP SIZE F/L CR OPTION FEM COMP SIZE F/R CR OPTION FEM COMP SIZE G/L CR OPTION FEM COMP SIZE G/R This device is indicated for patients with severe knee pain and disability

DateNov 29, 2017
StatusTerminated
ReasonThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Root causePackaging
ActionA similar recall was initiated in January 2016 to remove remaining inventory packaged in the old bag. Further evaluation identified additional lots packaged in the old bag that were not included in the January 2016 recall; therefore, resulting in this new recall for the additional lots packaged in the old bag. A firm, Zimmer Biomet, sent an "Urgent Medical Device Recall" notices dated November 29, 2017 and response forms to customers. The notice described the product, problem and actions to be taken. The Customers should take the following action: Risk Manager Responsibilities: 1. Review thi
Quantity13,227 in total
DistributionWorldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911572MG II POROUS COBAL T-CHROME FEMORAL COMPONENT
Zimmer, Inc., Jul 1, 1991
K933785NEXT GENERATION KNEE SYSTEM
Zimmer, Inc., Jan 30, 1995