ClearanceBrief
Recall Z-1181-2018 ยท FOX Stand, Infusion

Ohmeda Medical

Rotating IV Pole Rotating IV Pole falls under the category classified as capital equipment, an optional accessory that may be purchased to facilitate attachment and consolidation of patient equipment to one central patient care area. Devices commonly attached to IV poles include IV pumps, feeding pumps, and infusion fluid bags. Products used on: Giraffe OmniBed, Giraffe Warmers, Panda Warmers an

DateDec 1, 2017
StatusTerminated
ReasonThere is the potential for the Rotating IV Pole to fall. A fall could result in an injury to a bystander holding the patient or to the operator.
Root causeDevice Design
ActionOn December 1, 2017 an URGENT MEDICAL DEVICE CORRECTION letter was issued to customers notifying of the recall. The letter requested customers to discontinue use of the affected Rotating IV poles and remove from Service. Place affected Rotating IV poles in a quarantine marked area. A GE Service Representative will contact you to arrange for the removal & replacement of a new IV pole. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. For further questions, please call (410) 888-5376.
Quantity1,179 (1,004 US; 175 OUS)
DistributionUSA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.