Recall Z-1279-2018 ยท PQP Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System

Life Technologies Corporation

Oncomine Dx Target Test User Guides and Assay Definition File, Model: A32461; UDI: (01)10190302006071(11)170911(10)1.8(240)A32461 The Oncomine Dx Target Test is a qualitative in vitro diagnostic test that uses targeted high- throughput, parallel-sequencing technology to detect single-nucleotide variants (SNVs) and deletions in 23 genes from DNA and fusions in ROS1 from RNA isolated from formalin

DateDec 4, 2017
StatusTerminated
ReasonThe Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.
Root causeDevice Design
ActionThe following actions are described in the consignee later: Review your testing results from the Oncomine Dx Target Test to ensure that none of the reports containing BRAF V600E positive samples are impacted. Utilize Figure 1 and 2 to aid in identifying this discrepancy. If you have any questions, please contact 408.802.2341.
Quantity9
DistributionUS Distribution to the states of: AZ, CA, NC, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.