ClearanceBrief
Recall Z-1120-2018 ยท NFK Kit, Repair, Catheter, Hemodialysis

Medical Components, Inc dba MedComp

12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

DateFeb 23, 2018
StatusTerminated
ReasonThere is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Root causeNonconforming Material/Component
ActionOn February 23, 2018, Medcomp distributed Product Recall and Distribution notices to international distributors via email and to U.S. facilities via courier service on February 26, 2018. Medcomp is requesting the return of all un-used affected product. Customers should immediately examine inventory and quarantine product subject to recall. If for any reason the product was further distributed, please identify the recipients and notify them at once of this product recall. Your notification may be enhanced by including a copy of this recall notification letter. Contact your customer servic
Quantity712 kits
DistributionWorldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Irela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022570MEDCOMP REPAIR KIT, MODEL ASPCRPK
Medcomp, Oct 31, 2002