ClearanceBrief
Recall Z-1304-2018 ยท ILQ System, Communication, Powered

Dynavox Systems Llc

Dynavox T10 Speech Generating Device ( including Batteries)

DateDec 14, 2017
StatusTerminated
ReasonThe battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
Root causeComponent design/selection
ActionOn February 26, 2018 an URGENT MEDICAL RECALL LETTER was issued to customers with the subject line listing the serial number specific to that customer along with the date of purchase. The letter informs customers of the issue and instructs customers to return product to one of their facilities to have the product repaired until an RA. Questions or concerns can be directed to 1-800-344-1778, Press 1 and then Press 2.
Quantity4615
DistributionNationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.