ClearanceBrief
Recall Z-1073-2018 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Smiths Medical ASD Inc.

PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)

DateMar 15, 2018 (initiated Dec 4, 2017)
StatusTerminated
ReasonCertain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
Root causeProcess design
ActionCustomers were notified via letter on about 12/04/2017. Instructions include to examine current inventory for affected product, collect and isolate any potentially affected product, determine the total number of potentially affected product in possession and complete and return the Urgent Medical Device Recall Response Form, return any affected product on hand, and notify customers if the affected products were further distributed.
Quantity1,077 units total
DistributionWorldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973862PORT-A-CATH LOW-PROFILE IMPLANTABLE ACCESS SYSTEMS
Sims Deltec, Inc., Dec 29, 1997
K974471PORT-A-CATH- II DUAL-LUMEN LOW PROFILE SYSTEM, PORT-A-CATH II DUAL LUMEN LOW PROFILE SYSTEM WITH INTRODUCER SET, PORT-A
Sims Deltec, Inc., Jan 22, 1998