Recall Z-1459-2018 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Emotion Duo (2003) Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

DateFeb 15, 2018
StatusTerminated
ReasonTo inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
Root causeSoftware design
ActionThe Center for Devices and Radiological Health (CDRH) understands that the CAP involving 1301 units will consist of the following: 1. You will contact customers and initiate the actions to correct the failure to comply issue. 2. You will provide a software update to correct tube current calculations by the CARE Dose4D algorithm for head scans. 3. You will follow-up with consignees to verify the effectiveness of the software update. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this
Quantity1301
DistributionUSA (nationally) Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall