Recall Z-1451-2018 ยท JAK System, X-Ray, Tomography, Computed
Siemens Medical Solutions USA, Inc
SOMATOM Spirit Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
| Date | Feb 15, 2018 |
| Status | Terminated |
| Reason | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure |
| Root cause | Software design |
| Action | The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 1301 units will consist of the following: 1. You will contact customers and initiate the actions to correct the failure to comply issue. 2. You will provide a software update to correct tube current calculations by the CARE Dose4D algorithm for head scans. 3. You will follow-up with consignees to verify the effectiveness of the software update. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this |
| Quantity | 1301 |
| Distribution | USA (nationally) Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K042328 | SOMATOM SPIRIT Siemens Medical Systems, Inc., Sep 14, 2004 |