Recall Z-1184-2018 ยท JIX Calibrator, Multi-Analyte Mixture

Siemens Healthcare Diagnostics, Inc.

Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the calibration of follicle stimulating hormone (FSH), luteinizing hormone (LH), prolactin (PRL), and estradiol (E2) methods on the Dimension Vista System.

DateMar 26, 2018 (initiated Nov 9, 2017)
StatusTerminated
ReasonAn increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.
Root causeUnder Investigation by firm
ActionThe firm, Siemens, issued an "Urgent Medical Device Correction (VC18-01.A.US)" letter dated 11/3/2017, to US customers on November 3, 2017 and an "Urgent Field Safety Notice (VC18-01.A.OUS)" to customers outside the US in November 2017. Customers were instructed to: " Discontinue use of and discard impacted LOCI 8 CAL lot 7CD076 " Calibrate the E2 assay using an alternate lot of LOCI 8 CAL. " Review the letter with their Medical Director. " Review their inventory of this product to determine theirs laboratory's replacement needs and to provide information to Siemens for reporting to the reg
Quantity989 (681 USA/308 OUS)
DistributionWorldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Slovakia, Spain and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101460LOCI 8 CALIBRATOR, MODEL KC 646
Siemens Healthcare Diagnostics, Jul 9, 2010